Title: | Extension of the Calibration Curve for the PGRA Facility in Petten |
Authors: | NIEVAART Victor Alexander; STECHER-RASMUSSEN Finn; SAUERWEIN Wolfgang; WITTIG A.; MOSS Raymond; APPELMAN K. |
Citation: | APPLIED RADIATION AND ISOTOPES vol. 67 no. 7-8S p. S362-S364 |
Publisher: | PERGAMON-ELSEVIER SCIENCE LTD |
Publication Year: | 2009 |
JRC N°: | JRC52763 |
ISSN: | 0969-8043 |
URI: | www.elsevier.com/locate/apradiso http://publications.jrc.ec.europa.eu/repository/handle/JRC52763 |
DOI: | 10.1016/j.apradiso.2009.03.064 |
Type: | Articles in periodicals and books |
Abstract: | At the boron neutron capture therapy (BNCT) facility in Petten, the Netherlands, 10B concentrations in biological materials are measured with the prompt gamma ray analyses facility that is calibrated using certified 10B solutions ranging from 0 to 210 ppm. For this study, newly certified 10B solutions ranging up to 1972ppm are added. MCNP simulations of the setup range to 5000 ppm. A second order polynomial (as already used) will fit 10B-concentrations less than 300 ppm. Above 300ppm a fitted third order polynomial is needed to describe the calibration curve accurately. |
JRC Directorate: | Energy, Transport and Climate |