Title: Evaluation of a portable nephelometer against the Tapered Element Oscillating Microbalance method for monitoring PM2.5
Authors: KARAGULIAN FEDERICOBELIS CLAUDIOLAGLER FriedrichBARBIERE MAURIZIOGERBOLES Michel
Citation: JOURNAL OF ENVIRONMENTAL MONITORING vol. 14 no. 8 p. 2145-2153
Publisher: ROYAL SOC CHEMISTRY
Publication Year: 2012
JRC N°: JRC66906
ISSN: 1464-0325
URI: http://pubs.rsc.org/en/content/articlelanding/2012/em/c2em30099k
http://publications.jrc.ec.europa.eu/repository/handle/JRC66906
DOI: 10.1039/C2EM30099K
Type: Articles in Journals
Abstract: Monitoring personal exposure to particle matter (PM2.5) in ambient air requires performing measurements using portable monitors. In this work, the portable nephelometer SidePak AM510 Personal Aerosol Monitor manufactured by TSI Inc. was evaluated against a Tapered Element Oscillating Microbalance (TEOM) equipped with a Filter Dynamics Measurements System (FDMS). Conventionally, the SidePak is calibrated with respect to the Arizona Road Test Dust and then multiplied by an environmental calibration factor to yield mass concentration. To adapt this calibration to specific field conditions, we present an implementation of this calibration by introducing a growing factor correction which takes into account relative humidity and the dry and wet portions of the refractive index estimated from TEOM-FDMS measurements. PM2.5 sampling with several SidePaks AM510 was carried out in background and rural sites in the Po Valley (Italy). Modeled SidePak data were plotted vs. reference TEOM-FDMS data which show a good agreement.
JRC Institute:Institute for Environment and Sustainability

Files in This Item:
There are no files associated with this item.


Items in repository are protected by copyright, with all rights reserved, unless otherwise indicated.