Title: Intercomparison of radionuclide calibrators in Belgian hospitals
Authors: STRUELENS L.SALDARRIAGA VARGAS C.RODRIGUEZ PEREZ S.BAETE K.POMME STEFAANPAEPEN JANVAN AMMEL RAF
Publisher: The Belgian Society of Nuclear Medicine
Publication Year: 2015
JRC N°: JRC100111
URI: WWW.BELNUC.BE
http://publications.jrc.ec.europa.eu/repository/handle/JRC100111
Type: Articles in periodicals and books
Abstract: The SCK•CEN conducted a proficiency test of the capacity of Belgian hospitals to determine activity of Tc-99m, In-111 and F-18 in vials and syringes using radionuclide calibrators. The reference values were provided by the European Commissions Joint Research Centre in Geel. The results are presented and discussed. I key result is that calibration factors for syringes with In-111 need to be improved.
JRC Directorate:Health, Consumers and Reference Materials

Files in This Item:
There are no files associated with this item.


Items in repository are protected by copyright, with all rights reserved, unless otherwise indicated.