Deconvolution of 238,239,240Pu conversion electron spectra measured with a silicon drift detector
Internal conversion electron (ICE) spectra of thin 238,239,240Pu sources, measured with a windowless Peltier-cooled silicon drift detector (SDD), were deconvoluted and relative ICE intensities were derived from the fitted peak areas. Corrections were made for energy dependence of the full-energy-peak counting efficiency, based on Monte Carlo simulations. A good agreement was found with the theoretically expected internal conversion coefficient (ICC) values calculated from the BrIcc database.
POMME Stefaan;
MAROULI Maria;
PAEPEN Jan;
MARKOVIC Nikola;
PÖLLÄNEN R;
2018-03-12
PERGAMON-ELSEVIER SCIENCE LTD
JRC106649
0969-8043,
https://publications.jrc.ec.europa.eu/repository/handle/JRC106649,
10.1016/j.apradiso.2017.08.033,
Additional supporting files
File name | Description | File type | |