CERTIFICATION REPORT The certification of the absorbed energy (150 J nominal) of Charpy V-notch reference test pieces for tests at 20 °C: ERM®-FA415ao
This certification report describes the processing and characterisation of ERM®-FA415ao, a batch of Charpy V-notch certified reference test pieces certified for the absorbed energy (KV). Sets of five of these test pieces are used for the verification of pendulum impact test machines according to ISO 148-2 (Metallic materials - Charpy pendulum impact test – Part 2: Verification of testing machines).
The absorbed energy (KV) is operationally defined and refers to the impact energy required to break a V-notched test piece of standardised dimensions, as defined in ISO 148-1. The certified value of ERM-FA415ao is made traceable to the SI, via the SI-traceable certified value of the master batch ERM®-FA415s, by testing samples pieces of ERM-FA415ao and ERM-FA415s under repeatability conditions on an impact pendulum verified and calibrated with SI-traceably calibrated tools. The certified value is valid only for strikers with a 2 mm tip radius. The certified value is valid at (20 ± 2) °C.
VANDERVEKEN Francine;
SEGHERS John;
2019-12-17
Publications Office of the European Union
JRC118674
978-92-76-12973-8 (online),
1831-9424 (online),
EUR 29968 EN,
OP KJ-NA-29968-EN-N (online),
https://publications.jrc.ec.europa.eu/repository/handle/JRC118674,
10.2760/695934 (online),