Title: CERTIFICATION REPORT: The certification of electrophoretic mobility/zeta potential of silica particles in aqueous solution: ERM-FD306/SRM 1993
Authors: RAMAYE YANNICKESTENS VIKRAMCHAROUD-GOT JEANMAZOUA STEPHANEAUCLAIR GUYCHO TAE JOONTOMAN BLAZAHACKLEY V. A.LINSINGER THOMAS
Publisher: Publications Office of the European Union
Publication Year: 2020
JRC N°: JRC122048
ISBN: 978-92-76-23775-4 (online)
ISSN: 1831-9424 (online)
Other Identifiers: EUR 30410 EN
OP KJ-NA-30410-EN-N (online)
URI: https://publications.jrc.ec.europa.eu/repository/handle/JRC122048
DOI: 10.2760/753191
Type: EUR - Scientific and Technical Research Reports
Abstract: This report describes the production of ERM-FD306/SRM 1993, silica particles suspended in a borate buffer, certified for electrophoretic mobility and zeta potential by electrophoretic light scattering (ELS). This material was produced following ISO 17034:2016 and is certified in accordance with ISO Guide 35:2017. The certified reference material (CRM) was jointly produced by the Directorate F - Health, Consumers and Reference Materials of the European Commission’s Joint Research Centre (JRC) in Geel (Belgium) and the US National Institute of Standards and Technology (NIST), Gaithersburg (USA). The CRM was produced from a buffer-modified and diluted commercial colloidal silica slurry. Between-ampoule homogeneity was quantified and stability during dispatch and storage were assessed in accordance with ISO Guide 35:2017. The minimum sample intake for the ELS method was determined from the results and information provided by the laboratories that participated in the interlaboratory comparison (ILC) exercise. The material was characterised by an interlaboratory comparison between laboratories of demonstrated competence and adhering to ISO/IEC 17025. Technically invalid results were removed but no outlier was eliminated solely on statistical grounds. Uncertainties of the certified values were calculated in accordance with the Guide to the Expression of Uncertainty in Measurement (GUM) and include uncertainties related to possible inhomogeneity, instability and characterisation. The material is intended for quality control and assessment of method performance. The method-defined certified values are regarded as reliable estimates of the true values and ERM-FD306/SRM 1993 can therefore be used for calibration purposes. The CRM is available in 25 mL pre-scored amber glass ampoules each containing approximately 25 mL of suspension.
JRC Directorate:Health, Consumers and Reference Materials

Files in This Item:
File Description SizeFormat 
final_report_ermfd306.pdf3.53 MBAdobe PDFView/Open


Items in repository are protected by copyright, with all rights reserved, unless otherwise indicated.