Time interval distribution of nuclear events in a digital spectrometer
This study compares the theoretical formula for time interval density distributions (TIDs) in counters with pileup rejection to TIDs measured using a digital spectrometer with trapezoid filtering. The TIDs were derived from successive time stamps of 60Co and 133Ba gamma rays detected in a HPGe spectrometer. The results demonstrate that the theoretical formula is applicable up to 40% count loss and can be used to retrospectively deduce the incident event rate.
SOREN Cheng;
PIERSON Bruce;
POMME Stefaan;
FLASKA Marek;
2024-03-25
ELSEVIER
JRC134680
0168-9002 (online),
https://www.sciencedirect.com/science/article/pii/S016890022400144X?via%3Dihub,
https://publications.jrc.ec.europa.eu/repository/handle/JRC134680,
10.1016/j.nima.2024.169218 (online),