header.html

An official website of the European Union How do you know?      
European Commission logo

handle.jsp

cover
A-TOF is a novel time-of-flight device under construction for high-resolution spectroscopy of alpha-particle emission energies originating from the natural decay of heavy nuclides. It detects the time difference between pulses generated by the alpha-particle in a fly-through start-timing detector and in a stop-timing detector at several meters distance. This paper discusses the design of the start-timing detector and expands on details of its construction, including the electrostatic mirror, electron detector, and fast signal amplifier. Estimates are made about the anticipated timing resolution. The aim is to improve the energy resolution well below the physical limit of planar ion-implanted silicon detectors.
2025-09-19
PERGAMON-ELSEVIER SCIENCE LTD
JRC142262
1872-9800 (online),   
https://www.sciencedirect.com/science/article/pii/S0969804325005238,    https://publications.jrc.ec.europa.eu/repository/handle/JRC142262,   
10.1016/j.apradiso.2025.112178 (online),   
NameCountryCityType
Datasets
IDTitlePublic URL
Dataset collections
IDAcronymTitlePublic URL
Scripts / source codes
DescriptionPublic URL
Additional supporting files
File nameDescriptionFile type 