Certification of Charpy V-notch Reference Test Pieces of 30 J Nominal Absorbed Energy, ERM-FA013ax
This certification report describes the processing and characterisation of ERM-FA013ax, a batch of Charpy V-notch certified reference test pieces. Sets of five of these test pieces are used for the verification of pendulum impact test machines according to EN 10045-2 or according to ISO 148-2. The certified value for KV (absorbed energy = energy required to break a V-notched test piece using a pendulum impact test machine) is 31.1 J. The associated expanded uncertainty (1.3 J, k = 2 corresponding to a confidence level of about 95 %) is calculated for the mean of a set of five test pieces. The certified value is traceable to the Charpy impact test method as described in EN 10045 1 (3) and ISO 148 (4), via the corresponding Master Batch ERM-FA013d of the same nominal absorbed energy (30 J).
ROEBBEN Gert;
LAMBERTY Andree;
2007-08-21
JRC37554
978-92-79-06106-6,
EUR 22785 EN,
OP LA-NA-22785-EN-C,
https://publications.jrc.ec.europa.eu/repository/handle/JRC37554,
Additional supporting files
| File name | Description | File type | |