Certification of Charpy V-notch Reference Test Pieces of 60 J Nominal Absorbed Energy, ERM®-FA014p
This certification report describes the processing and characterisation of ERM®-FA014p, a batch of Charpy V-notch certified reference test pieces. Sets of fieve of these test pieces are used for the verification of pendulum impact test machines according to EN 10045-2 or according to ISO 148-2. The certified value for KV (= energy required to break a V-notch test piece using a pendulum impact test machine) is 66.6 J. The associated uncertainty (1.6 J, k=2 corresponding to a confidence level of 95 %) is calculated for the mean of a set of five test pieces. The certified value is traceable to the Charpy impact test method as described in EN 10045 1 and ISO 148, via the corresponding Master Batch ERM®-FA014c of the same nominal absorbed energy (60 J).
ROEBBEN Gert;
LAMBERTY Andree;
2008-02-27
OPOCE
JRC43684
978-92-79-05126-5,
1018-5593,
EUR 23262 EN,
OP LA-NA-23262-EN-C,
https://publications.jrc.ec.europa.eu/repository/handle/JRC43684,
10.2787/31095,