Certifcation of Charpy V-Notch Reference Test Pieces of 80 J Nominal Absorbed Energy (ERM-FA015w)
This certification report describes the processing and characterisation of ERM®-FA015w, a batch of Charpy V-notch certified reference test pieces. Sets of five of these test pieces are used for the verification of pendulum impact test machines according to EN 10045-2 or according to ISO 148-2. The certified value for KV (= energy required to break a V-notched test piece using a pendulum impact test machine) is 81.2 J. The associated uncertainty (2.8 J, k = 2 corresponding to a confidence level of 95 %) is calculated for the mean of a set of five test pieces. The absorbed energy (KV) is procedurally defined in EN 10045-1 and ISO 148. The certified value of ERM®-FA015w is traceable to the SI, via the SI-traceable certified value of the master batch ERM®-FA015w, by testing samples of ERM®-FA015v and ERM®-FA015w in repeatability conditions on an impact pendulum verified with SI-traceably calibrated tools.
ROEBBEN Gert;
2010-03-17
Publications Office of the European Union
JRC57029
978-92-79-15222-1,
1018-5593,
EUR 24277 EN,
OP LA-NA-24277-EN-C,
https://publications.jrc.ec.europa.eu/repository/handle/JRC57029,
10.2787/24334,
Additional supporting files
| File name | Description | File type | |