Full metadata record
DC FieldValueLanguage
dc.contributor.authorROEBBEN Gerten_GB
dc.contributor.authorDEAN Alanen_GB
dc.contributor.authorLAMBERTY MARIE ANDREEen_GB
dc.date.accessioned2013-10-10T00:03:08Z-
dc.date.available2013-10-09en_GB
dc.date.available2013-10-10T00:03:08Z-
dc.date.created2013-09-05en_GB
dc.date.issued2013en_GB
dc.date.submitted2013-08-27en_GB
dc.identifier.isbn978-92-79-32688-2 (pdf)en_GB
dc.identifier.issn1831-9424 (online)en_GB
dc.identifier.otherEUR 26109en_GB
dc.identifier.otherOPOCE LA-NA-26109-EN-N (online)en_GB
dc.identifier.urihttp://publications.jrc.ec.europa.eu/repository/handle/JRC84079-
dc.description.abstractThis certification report describes the processing and characterisation of ERM®-FA415aa, a batch of Charpy V-notch certified reference test pieces certified for the absorbed energy (KV) for tests at 20 °C. Sets of five of these test pieces are used for the verification of pendulum impact test machines according to ISO 148-2 (Metallic materials - Charpy pendulum impact test – Part 2: Verification of testing machines). The absorbed energy (KV) is procedurally defined and refers to the impact energy required to break a V-notched test piece of standardised dimensions, as defined in ISO 148-1. The certified value of ERM- FA415aa is traceable to the SI, via the SI-traceable certified value of the master batch ERM-FA415s, by testing samples of ERM-FA415aa and ERM-FA415s under repeatability conditions on an impact pendulum verified and calibrated with SI-traceably calibrated tools. The certified value is valid only for strikers with a 2 mm tip radius. The certified values are valid at (20 ± 2) °C.en_GB
dc.description.sponsorshipJRC.D.2-Standards for Innovation and sustainable Developmenten_GB
dc.format.mediumOnlineen_GB
dc.languageENGen_GB
dc.publisherPublications Office of the European Unionen_GB
dc.relation.ispartofseriesJRC84079en_GB
dc.titleThe certification of the absorbed energy (150 J nominal) of Charpy V-notch reference test pieces for tests at 20 °C: ERM®-FA415aaen_GB
dc.typeEUR - Scientific and Technical Research Reportsen_GB
dc.identifier.doi10.2787/6808 (online)en_GB
JRC Directorate:Health, Consumers and Reference Materials

Files in This Item:
File Description SizeFormat 
report erm-fa415aa.pdf1.06 MBAdobe PDFView/Open


Items in repository are protected by copyright, with all rights reserved, unless otherwise indicated.