Report of laboratory and in-situ validation of micro-sensor for monitoring ambient air pollution - NO9: CairClipNO2 of CAIRPOL (F)
The aim of this report is to evaluate and validate CairClipNO2 sensors of CAIRPOL within laboratory and field tests under ambient/indoor air conditions corresponding to a specific micro-environment: traffic station, urban/suburban areas. This report presents the evaluation of the performances and determination of the the laboratory and field measurement uncertainty of the sensor values, compared to uncertainties fixed by the Data Quality Objective (DQO) of the European Air Quality Directive for indicative method. Further, procedures for the calibration of sensors able to ensure full traceability of measurements of sensors to SI units is developed.
SPINELLE Laurent;
GERBOLES Michel;
ALEIXANDRE Manuel;
2014-04-24
Publications Office of the European Union
JRC86499
978-92-79-34920-1,
1831-9424,
EUR 26394,
OP LB-NA-26394-EN-N,
https://publications.jrc.ec.europa.eu/repository/handle/JRC86499,
10.2788/4975,
Additional supporting files
| File name | Description | File type | |