Excess heat capacity of the (Li1-x, Cax )F1+x liquid solution determined by Differential Scanning Calorimetry and drop calorimetry
The work presents the measured heat capacity of the (Li,Ca)Fx liquid solution. Four samples with different compositions have been prepared and measured using a Differential Scanning Calorimeter. Since this technique was newly adopted for measuring encapsulated fluoride samples, some modifications were introduced in the standard configuration of the instrument and they are described in this work as well. For comparison one of the analysed composition (xCaF2 = 0.5) was also measured using drop calorimetry, which has been previously used for similar studies. The reliability of the results obtained was confirmed by the good agreement between the two techniques. Moreover, the excess heat capacity of the (Li,Ca)Fx liquid solution was derived and a strong deviation from the ideal behaviour was observed.
CAPELLI Elisa;
BENES Ondrej;
KONINGS Rudy;
2014-12-11
ACADEMIC PRESS LTD- ELSEVIER SCIENCE LTD
JRC90914
0021-9614,
http://www.sciencedirect.com/science/article/pii/S0021961414003267,
https://publications.jrc.ec.europa.eu/repository/handle/JRC90914,
10.1016/j.jct.2014.10.007,
Additional supporting files