A comparison was conducted, between 11 European National Metrology Institutes and EC-JRC, on measurement of 137Cs activity concentration in fume dust. As comparison material an activity standard produced from real contaminated fume dust was used. The standard material consisted of 13 samples of compressed fume dust, each of 50 g mass and of well-defined cylindrical geometry (diameter 69.5 mm and height 18.7 mm). The material contained 137Cs and 60Co of reference activity concentrations of (9.72 ± 0.10) Bq/g and (0.450 ± 0.018) Bq/g, respectively, determined using the comparison results. The organisation and results of the intercomparison, as well as the process of obtaining reliable reference values are presented.
TZIKA Faidra;
HULT Mikael;
BURDA O.;
ARNOLD D.;
SIBBENS Goedele;
CARO MARROYO B.;
GOMEZ-MANCEBO M. B.;
PEYRES Virginia;
MOSER H.;
FERREUX L.;
SOLC J.;
DRYAK P.;
FAZIO A.;
LUCA A.;
VODENIK Branko;
REIS M.;
TYMINSKI Z.;
KLEMOLA S.;
2016-01-04
PERGAMON-ELSEVIER SCIENCE LTD
JRC92124
0969-806X,
http://www.journals.elsevier.com/radiation-physics-and-chemistry/open-access-articles/,
https://publications.jrc.ec.europa.eu/repository/handle/JRC92124,
10.1016/j.radphyschem.2015.03.043,
This document is only visible at the Commission level.
You are not authorized to publish or distribute it outside the European Commission.
This is a public document. You can share this publication.
Additional supporting files
| File name | Description | File type | |