Title: Performance evaluation of amperometric sensors for the monitoring of O3 and NO2 in ambient air at ppb level
Authors: SPINELLE LAURENTGERBOLES MichelALEIXANDRE Manuel
Citation: PROCEDIA ENGINEERING vol. 120 p. 480-483
Publisher: ELSEVIER BV
Publication Year: 2015
JRC N°: JRC96920
ISSN: 1877-7058
URI: http://www.sciencedirect.com/science/article/pii/S1877705815023395
http://publications.jrc.ec.europa.eu/repository/handle/JRC96920
DOI: 10.1016/j.proeng.2015.08.676
Type: Articles in periodicals and books
Abstract: Hereafter we present a laboratory evaluation of commercially available amperometric sensors for the monitoring of O3 and NO2 in ambient air at ppb level. The tests includes the determination of the response time of sensors, their calibration function, the evaluations of repeatability, short and long term drifts, the hysteresis effect and the matrix effect. Interferences from gaseous compounds, temperature and humidity parameters are also evaluated.
JRC Directorate:Sustainable Resources

Files in This Item:
There are no files associated with this item.


Items in repository are protected by copyright, with all rights reserved, unless otherwise indicated.