The Recoil Proton Telescope in Non-coincidence Mode for Neutron Fluence Measurements
The paper describes an alternative approach to use the Los Alamos type Recoil Proton Telescope for fluence measurements of mono-energetic neutrons with energies above 1 MeV. Traditionally, this type of detectors are used in a 3-coincidence acquisition mode (proportional counter – proportional counter – end-detector). This study shows that by only using the end-detector and a background reduction method similar results are obtained, but with considerably less complexity in the measurement set-up.
LOEVESTAM Goeran;
2006-12-05
ELSEVIER SCIENCE BV
JRC33340
https://publications.jrc.ec.europa.eu/repository/handle/JRC33340,
Additional supporting files
| File name | Description | File type | |