Within the harmonization program of Air Quality monitoring in Europe ERLAP Laboratories are organizing Inter-Laboratory Comparison in the facility of Ispra (Italy). From the 30th of September to the 3rd of October 2013 in Ispra (IT), eight Laboratories of AQUILA (Network of European Air Quality Reference Laboratories) met for a laboratory comparison exercise to evaluate their proficiency in the analysis of inorganic gaseous pollutants covered by European Directive about air quality (SO2, CO, NO, NO2 and O3).
The proficiency evaluation, where each participant’s bias was compared to two criteria, provides information on the current situation and capabilities to the European Commission and can be used by participants in their quality control system.
On the basis of criteria imposed by the European Commission, 89.4% of the results reported by AQUILA laboratories were good both in terms of measured values and reported uncertainties. The rest 10.6% of the results had good measured values, but the reported uncertainties were either too high (7.3%) or too small (3.3%).
Comparability of results among AQUILA participants at the highest concentration level, excluding outliers, is acceptable in for all pollutants measurements.
LAGLER Friedrich;
BARBIERE Maurizio;
2014-05-02
Publications Office of the European Union
JRC89757
978-92-79-37796-9,
1831-9424,
EUR 26604,
OP LB-NA-26604-EN-N,
https://publications.jrc.ec.europa.eu/repository/handle/JRC89757,
10.2788/52182,
| Name | Country | City | Type |
|---|
This document is only visible at the Commission level.
You are not authorized to publish or distribute it outside the European Commission.
This is a public document. You can share this publication.